KB-112 · IP-8100 infusion pump
Downstream occlusion alarm with a clear line
- 01Run the occlusion test on a calibrated test set at 25 mL/h.
- 02Read the downstream pressure in service mode. Above 12 psi with no occlusion points to the sensor.
- 03Replace pressure sensor assembly 81-2210, then recalibrate.
- 04Electrical safety test before return to service.